|
Indications, Contraindications, Warnings, Precautions,
and adverse effects
Torqr®
6F
Intended Use
The Medtronic CardioRhythm Torqr Catheter is intended for use in
diagnostic electrophysiologic procedures. The catheter is designed
for recording intracardiac electrograms and temporary pacing associated
with electrophysiology studies.
Contraindications
There are no known contraindications for this catheter.
Warnings
- This device is for single use only. Do not resterilize and/or
reuse.
- Catheter resterilization and reuse may result in loss of proper
electrical or mechanical function and could cause patient injury.
- This catheter should be used by or under the supervision of
physicians thoroughly trained in electrophysiology, including
the placement and use of intracardiac electrode catheters.
- Electrical recording or stimulation equipment must be front
end isolated, or have an isolated patient cable. Current
leakage from any connected electrical equipment must not exceed
10 µA for intracardiac electrodes.
- Perforation of the vasculature is an inherent risk of
any catheter placement. Additional potential complications are
those attending any intracardiac catheterization procedure including,
but not limited to: cardiac tamponade, thromboembolic episodes,
hematoma, pneumothorax, local or systemic infection.
- US LAW DOES NOT PERMIT THE USE OF THIS CATHETER FOR INTRACARDIAC
ABLATION. DO NOT USE FOR DC ABLATION. PRECAUTIONS
- To maintain optimal patient safety and catheter electrode integrity
do not wipe this catheter with organic solvents, such as alcohol.
- Excessive bending or kinking of catheter may damage internal
electrode wires and/or distal tip shaping capabilities.
- Catheter materials are not compatible with magnetic resonance
imaging (MRI).
return to top
Torqr®
5F
Indications for use
The Medtronic CardioRhythm Torqr Catheter is intended for use in
diagnostic electrophysiologic procedures. The catheter is designed
for recording intracardiac electrograms and temporary pacing associated
with electrophysiology studies.
Contraindications
There are no known contraindications for this catheter.
Warnings and precautions
- This device is for single use only. Do not resterilize and/or
reuse. Catheter resterilization and reuse may result in loss of
proper electrical or
mechanical function and could cause patient injury.
- This catheter should be used by or under the supervision of
physicians
thoroughly trained in electrophysiology, including the placement
and use of
intracardiac electrode catheters.
- Electrical recording or stimulation equipment must be front
end isolated,
or have an isolated patient cable. Current leakage from any connected
electrical equipment must not exceed 10 µA for intracardiac
electrodes.
- Perforation of the vasculature is an inherent risk of any catheter
placement. Additional potential complications are those attending
any
intracardiac catheterization procedure including, but not limited
to: cardiac
tamponade, thromboembolic episodes, hematoma, pneumothorax, local
or
systemic infection.
- US LAW DOES NOT PERMIT THE USE OF THIS CATHETER FOR
INTRACARDIAC ABLATION. DO NOT USE FOR DC ABLATION.
Precautions
- To maintain optimal patient safety and catheter electrode integrity
do not
wipe this catheter with organic solvents, such as alcohol.
- Excessive bending or kinking of catheter may damage internal
electrode
wires and/or distal tip shaping capabilities.
- Catheter materials are not compatible with magnetic resonance
imaging
(MRI).
return to top
Soloist
Intended Use
The Medtronic CardioRhythm Soloist catheter is intended for use
in diagnostic electrophysiologic procedures. The catheter is designed
for recording intracardiac electrograms and temporary pacing associated
with electrophysiology studies.
Contraindications
There are no known contraindications for this catheter.
Warnings
- This device is for single use only. Do not resterilize and/or
reuse. Catheter resterilization and reuse may result in loss of
proper electrical or mechanical function and could cause patient
injury.
- This catheter should be used by or under the supervision of
physicians thoroughly trained in electrophysiology, including
the placement and use of intracardiac electrode catheters.
- Electrical recording or stimulation equipment must be "front
end" isolated, or have an isolated patient cable. Current
leakage from any connected electrical equipment must not exceed
10 µA for intracardiac electrodes.
- Do not use in the coronary vasculature other than in the coronary
sinus.
- Perforation of the vasculature is an inherent risk of
any catheter placement. Additional potential complications are
those attending any intracardiac catheterization procedure including,
but not limited to: cardiac tamponade, thromboembolic episodes,
hematoma, pneumothorax, local or systemic infection.
- US LAW DOES NOT PERMIT THE USE OF THIS CATHETER FOR INTRACARDIAC
ABLATION. DO NOT USE FOR DC ABLATION.
Precautions
- To maintain optimal patient safety and catheter electrode integrity
do not wipe this catheter with organic solvents, such as alcohol.
- Excessive bending or kinking of catheter may damage internal
electrode wires.
- Catheter materials are not compatible with magnetic resonance
imaging (MRI).
return to top
Stablemapr
SM
Intended Use
The Medtronic StableMapr catheter is intended for use in diagnostic
electrophysiologic procedures. These catheters are designed for
recording intracardiac electrograms and temporary pacing associated
with electrophysiology studies.
Contraindications
There are no known contraindications for these catheters. 2 English
Instructions for use
Warnings
- This device is for single use only. Do not resterilize and reuse
the catheter. This may result in loss of proper electrical and
mechanical function.
- This catheter should be used only by or under the supervision
of physicians well trained in electrophysiology, including the
placement and use of intracardiac electrode catheters.
- Electrical recording or stimulation equipment must be front
end isolated, or have an isolated patient cable (IEC 601-1
Type CF equipment or equivalent). Current leakage from any connected
electrical equipment must not exceed 10 µA for intracardiac
electrodes.
- Perforation of the vasculature is an inherent risk of any catheter
placement. Additional potential complications are those attending
any intracardiac catheterization procedure including, but not
limited
to: cardiac tamponade, thromboembolic episodes, hematoma, pneumothorax,
local or systemic infection and death.
- Use only connecting cables supplied by Medtronic or patient
injury, operator injury, or equipment damage may occur.
- Do not allow moisture into connectors of the StableMapr catheter
or connecting cables. If the connectors get wet, the system may
not function correctly.
- The safety and effectiveness of this device as an ablation catheter
have not been established. Therefore, such use is considered investigational.
Precautions
- To maintain optimal patient safety and catheter electrode integrity
do not wipe this catheter with organic solvents, such as alcohol.
- Excessive bending or kinking of catheter may damage internal
electrode wires and/or distal tip shaping capabilities.
- Catheter materials are not compatible with magnetic resonance
imaging (MRI).
return to top
RF
Conductr® MC
Intended Use
The RF Conductr MC catheter is intended for use with the Atakr RF
Power Generator to deliver RF energy for intracardiac ablation of
accessory atrioventricular (AV) conduction pathways associated with
tachycardia, for the treatment of AV nodal re-entrant tachycardia,
and for creation of complete AV block in patients with a difficult
to control ventricular response to an atrial arrhythmia.
Contraindications
The use of this device is contraindicated in patients with active
systemic infection.
The transseptal approach is contraindicated in patients with left
atrial thrombus or myxoma, or interatrial baffle or patch. The retrograde
transaortic approach is contraindicated in patients with aortic
valve replacement.
Warnings
- Catheter ablation procedures present the potential for significant
x-ray exposure, which can result in acute radiation injury
as well as increased risk for somatic and genetic effects, to
both patients and laboratory staff due to the x-ray beam intensity
and duration of the fluoroscopic imaging. Catheter ablation should
only be performed after adequate attention has been given to the
potential radiation exposure associated with the procedure, and
steps taken to minimize this exposure. Careful consideration must
therefore be given to the use of the device in pregnant women.
- During a transaortic approach, adequate fluoroscopic visualization
is necessary to avoid placement of the ablation catheter within
the coronary vasculature. Catheter placement and RF power application
within a coronary artery has been associated with myocardial infarction
and death.
- Catheter materials are not compatible with magnetic resonance
imaging (MRI).
- Perforation of the vasculature is an inherent risk of
any catheter placement. A number of serious adverse events have
been documented for catheter ablation procedures including pulmonary
embolism, myocardial
infarction, stroke, cardiac tamponade, and death. See Adverse
Events, for additional potential complications.
- Patients undergoing left-sided ablation procedures should
be closely monitored during the post-ablation period for clinical
manifestations of infarction.
- Catheters with distal pair electrode spacing greater than
2 mm should not be used in the ablation of septal accessory
pathways or in the treatment of AV nodal re-entrant tachycardia
because of the potential for creating inadvertent complete AV
block.
- Implantable pacemakers and implantable cardioverter/defibrillators
(ICDs) may be adversely affected by radiofrequency (RF) current.
It is important to: have temporary external sources of pacing
and defibrillation available during ablation; exercise extreme
caution during ablation when in close proximity to atrial or ventricular
permanent leads; and, perform complete implantable device analysis
for all patients after ablation.
- Implanted cardioverter/defibrillators should be deactivated
during delivery of RF power.
- Patients undergoing AV node modification or septal accessory
pathway ablation are at risk for complete AV block. Permanent
pacing was required in 1.6% (2/128) of septal accessory pathway,
3.1% (1/32) of left posterior pathways, and 1.7% (4/238) of AV
nodal modification patients who experienced inadvertent partial
or complete AV block during the study. Closely monitor AV conduction
during RF energy delivery. Immediately terminate energy delivery
if partial or complete AV block is noted.
- The Atakr RF Power Generator is capable of delivering significant
RF power. Do not touch the distal tip of the CardioRhythm
RF catheter and the dispersive electrode at the same time especially
while operating the Atakr RF Power Generator or operator injury
may occur.
- Use only isolated amplifiers, pacing equipment, and ECG equipment
(IEC 601-1 Type CF equipment, or equivalent) or patient injury
or death may occur. Leakage current from any connected device
to the patient must not exceed 10 micro Amps (µA) under
any circumstances.
- The CardioRhythm RF catheters are disposable and intended
for single use only. Do not resterilize or reuse. This may
result in loss of proper electrical and mechanical function and
could cause patient injury and/or the
communication of infectious diseases from one patient to another.
- Precautions
Do not attempt to operate the Atakr Ablation System or connect
the catheter to the Atakr RF Power Generator prior to completely
reading and understanding the Atakr Ablation System Technical
Manual and the RF Conductr MC catheter Instructions For Use.
- Cardiac ablation procedures should be performed only by appropriately
trained personnel in a fully-equipped electrophysiology
laboratory.
- This catheter should be used only by or under the supervision
of physicians well trained in electrophysiology, including the
placement and use of intracardiac electrode catheters and experienced
in performing radiofrequency catheter ablation procedures.
- The long-term risks of protracted fluoroscopy have not
been established.
Careful consideration must therefore be given for the use of the
device in prepubescent children.
- The long-term risks of lesions created by RF ablation have
not been established. In particular, any long-term effects
of lesions in proximity to the specialized conduction system or
coronary vasculature are unknown.
Furthermore, the risk/benefit in asymptomatic patients has not
been studied.
- The CardioRhythm RF catheters are intended for use only with
the Atakr RF Power Generator and accessories. The safety and
use with other RF power generators or accessories has not been
tested. Use only
Medtronic CardioRhythm cables supplied with the Atakr Ablation
System.
- Excessive bending or kinking of catheter may damage internal
electrode wires and/or distal tip shaping capabilities.
- Do not allow moisture into connectors on the CardioRhythm
RF catheter, Atakr RF Power Generator, or cables. If the connectors
get wet, the system may not function correctly.
- The sterile packaging and catheter should be inspected prior
to use. If damaged, do not use; contact your local Medtronic representative.
- Careful catheter manipulation must be performed in order to
avoid cardiac damage, perforation or tamponade. Catheter advancement
should be performed under fluoroscopic guidance. Do not
use excessive force to advance or withdraw the catheter when resistance
is encountered.
- See Atakr Technical Manual (Section 9.10) for information concerning
catheter removal following generator shutdown.
- Do not deliver DC energy through the Atakr Ablation System
or the CardioRhythm RF catheter. Neither the Atakr Ablation System
nor the CardioRhythm RF catheter were designed to deliver DC energy
and no testing has been performed.
- Verify that RF catheter tip electrode is not in contact with
electrodes of other catheters in the heart or heating of the
other electrodes may occur.
- The catheter impedance display should be monitored during
energy delivery. If a sudden impedance rise is observed during
the ablation procedure, energy delivery should be discontinued
and the catheter tip examined for coagulum, and coagulum removed
if present.
- Do not begin an ablation procedure if the LOW BATTERY
display is illuminated on the Atakr RF Power Generator. Replace
battery.
- To maintain optimal patient safety and catheter electrode integrity
do not wipe this catheter with organic solvents, such as
alcohol.
- Bench testing of CardioRhythm RF catheters demonstrated that
they could withstand twenty-five RF energy deliveries without
any reduction in performance. During the clinical trial as many
as 73 energy deliveries were made with a single catheter, although
the average number was about seven.
Clinical Summary
Please refer to the Atakr Ablation System Technical Manual for the
Clinical Background.
Adverse Events
One or more adverse events were reported in 67 of the 683 patients
(a total of 88 adverse events) during the clinical study of the
Atakr Ablation System. The following adverse events are listed in
descending order according to their clinical significance as determined
by their severity and frequency
(< 1% unless otherwise indicated).
Death |
Valvular
Insufficiency |
Pericarditis |
| Cardiac
Tamponade |
Femoral
Artery |
Pleural
Effusion |
| Cerebrovascular |
Laceration
|
Respiratory
Depression |
|
Accident
|
Thrombus/Embolic |
Atrial
Lead |
| Myocardial
Infarction |
Event
|
Dislodgement
|
| Unintended
AV or |
Bowel
Obstruction |
Infected
IV Site |
|
Bundle
Branch
|
Brachial
Plexus Injury |
Hypotension |
|
Block
Requiring new
|
Pneumonia |
Bradycardia |
|
pacemaker
(1.0%)
|
Inferior
Vena |
Vasovagal
Reaction |
| Coronary
Sinus |
Cava
Clot
|
Thrombophlebitis |
|
Perforation
|
Deep Vein |
Temperature
Elevation |
| Bacterial
Endocarditis |
Thrombosis
|
Puncture
Site |
| Ventricular
Fibrillation |
Pericardial |
Hematoma
(1.6%)
|
| Coronary
Artery Spasm |
Effusion
(2.2%)
|
Phlebothrombosis |
|
|
High CPK |
Other complications of catheter ablation
procedures which have been reported in the literature include:
| Pulmonary
Embolism |
Nerve
or Blood |
Transient
Ischemic Attack |
| Thromboembolism |
Vessel
Injury
|
Infection |
There were 18 deaths reported among the
683 patients enrolled in the study. None of the 18 deaths were judged
by the investigators as definitely related to the use of the Atakr
RFCA System, but two deaths were judged as possibly device related
and a third as possibly procedure related.
Device related complications were reported
in 30 of the 683 patients (4%) including unintentional AV block
in 13 patients (1.9%), pericardial effusion in 8 patients (1.2%),
and cardiac tamponade in 3 patients (0.4%).
return to top
5mm
Tip RF Conductr®
Intended use
The RF Conductr MC - 5 mm Tip catheter is intended for use with
a Medtronic RF power generator to deliver RF energy for intracardiac
ablation of accessory atrioventricular (AV) conduction pathways
associated with tachycardia, for the treatment of AV nodal re-entrant
tachycardia, and for creation of complete AV block in patients with
a difficult to control ventricular response to an atrial arrhythmia.
Contraindications
The use of this device is contraindicated in patients with active
systemic infection.
The transeptal approach is contraindicated in patients with left
atrial thrombus or myxoma, or interatrial baffle or patch.
The retrograde transaortic approach is contraindicated in patients
with aortic valve replacement.
Warnings
- Catheter ablation procedures present the potential for significant
x-ray exposure, which can result in acute radiation injury
as well as increased risk for somatic and genetic defects, to
both patients and laboratory staff due to the x-ray beam intensity
and duration of the fluoroscopic imaging. Catheter ablation should
only be performed after adequate attention has been given to the
potential radiation exposure associated with the procedure, and
steps taken to minimize this exposure. Careful consideration must
therefore be given to the
use of the device in pregnant women.
- During a transaortic approach, adequate fluoroscopic visualization
is necessary to avoid placement of the ablation catheter within
the coronary vasculature. Catheter placement and RF power application
within a coronary artery has been associated with myocardial infarction
and death.
- Catheter materials are not compatible with magnetic resonance
imaging (MRI).
- Perforation of the vasculature is an inherent risk of
any catheter placement. A number of serious adverse events have
been documented for catheter ablation procedures including pulmonary
embolism, myocardial infarction, stroke, cardiac tamponade, and
death. See Adverse events for additional potential complications.
- Patients undergoing left-sided ablation procedures should
be closely monitored during the post-ablation period for clinical
manifestations of infarction.
- Catheters with distal pair electrode spacing greater than
2 mm should not be used in the ablation of septal accessory
pathways or in the treatment of AV nodal re-entrant tachycardia
because of the potential for creating inadvertent complete AV
block.
- Implantable pacemakers and implantable cardioverter/defibrillators
(ICDs) may be adversely affected by RF current. It is important
to: have temporary external sources of pacing and defibrillation
available during ablation; exercise extreme caution during ablation
when in close proximity to atrial or ventricular permanent leads;
and perform complete implantable device analysis for all patients
after ablation.
- ICDs should be deactivated during delivery of RF power.
- Patients undergoing AV node modification or septal accessory
pathway ablation are at risk for complete AV block. Permanent
pacing was required in 1.6% (2/128) of septal accessory pathway,
3.1% (1/32) of left posterior pathways, and 1.7% (4/238) of AV
nodal modification patients who experienced inadvertent partial
or complete AV block during a Medtronic clinical study. Closely
monitor AV conduction during RF energy delivery. Immediately terminate
energy delivery if partial or complete AV block is noted. Medtronic
RF power generators are capable of delivering significant RF power.
Do not touch the distal tip or the adjacent shaft electrode(s)
of the RF Conductr MC - 5 mm Tip catheter and the dispersive electrode
at the same time, especially while operating the
- Medtronic RF power generator, or operator injury may occur.
Use only isolated amplifiers, pacing equipment, and ECG equipment
(IEC 601-1 Type CF equipment, or equivalent) or patient injury
or
death may occur. Leakage current from any connected device to
the patient must not exceed 10 micro Amps (mA) under any circumstances.
- The RF Conductr MC - 5 mm Tip catheters are disposable and
intended for single use only. Do not resterilize or reuse.
This may result in loss of proper electrical and mechanical function
and could
cause patient injury and/or the communication of infectious diseases
from one patient to another.
Precautions
- Do not attempt to operate the Medtronic ablation system or connect
the RF Conductr MC - 5 mm Tip catheter to a Medtronic RF power
generator prior to completely reading and understanding the
- Medtronic ablation system technical manual and the RF Conductr
MC - 5 mm Tip catheter instructions for use.
- Cardiac ablation procedures should be performed only by appropriately
trained personnel in a fully equipped electrophysiology
laboratory.
- This catheter should only be used by or under the supervision
of physicians well trained in electrophysiology, including the
placement and use of intracardiac electrode catheters, and experienced
in performing RF catheter ablation procedures.
- The long term risks of protracted fluoroscopy have not
been established. Careful consideration must therefore be given
for the use of the device in prepubescent children.
- The long term risks of lesions created by RF ablation have
not been established. In particular, any long term effects
of lesions in proximity to the specialized conduction system or
coronary vasculature are unknown. Furthermore, the risk/benefit
in asymptomatic patients has not been studied.
- The RF Conductr MC - 5 mm Tip catheters are intended for use
only with a Medtronic RF power generator and accessories.
The safety and use with other RF power generators or accessories
has not been tested. Use only Medtronic cables.
- Excessive bending or kinking of the catheter may damage
internal electrode wires and/or distal tip shaping capabilities.
- Do not allow moisture into the connectors on the RF Conductr
MC - 5 mm Tip catheter, Medtronic RF power generator, or cables.
If the connectors get wet, the system may not function correctly.
- The sterile packaging and catheter should be inspected prior
to use. If damaged, do not use; contact your local Medtronic representative.
- Careful catheter manipulation must be performed in order to
avoid cardiac damage, perforation, or tamponade. Catheter advancement
should be performed under fluoroscopic guidance. Do not
use
excessive force to advance or withdraw the catheter when resistance
is encountered.
- See the Medtronic ablation system technical manual for information
concerning catheter removal following generator shutdown.
- 4 English Instructions for use
Do not deliver DC energy through the Medtronic ablation
system or the RF Conductr MC - 5 mm Tip catheter. Neither the
Medtronic ablation system nor the RF Conductr MC - 5 mm Tip catheter
were designed to deliver DC energy and no testing has been performed.
- Verify that the RF Conductr MC - 5 mm Tip catheter tip electrode
is not in contact with electrodes of other catheters in the
heart or heating of the other electrodes may occur during RF energy
delivery.
- The catheter impedance display should be monitored during
energy delivery. If a sudden impedance rise is observed during
the ablation procedure, energy delivery should be discontinued,
the catheter ablation electrodes examined, and coagulum removed
if present.
- Do not begin an ablation procedure if the LOW BATTERY
display is illuminated on the Medtronic RF power generator. Replace
battery.
- To maintain optimal patient safety and catheter electrode integrity
do not wipe the catheter with organic solvents, such as
alcohol.
- Bench testing of Medtronic RF catheters demonstrated that they
could withstand twenty-five RF energy deliveries without any reduction
in performance.
Clinical summary
Please refer to the Medtronic ablation system technical manual for
the clinical background.
Adverse events
One or more adverse events were reported in 67 of the 683 patients
(a total of 88 adverse events) during the clinical study of the
Medtronic ablation system. The following adverse events are listed
in descending order
according to their clinical significance as determined by their
severity and frequency (<1% unless otherwise indicated). Other
complications of catheter ablation procedures which have been reported
in the literature include:
| Death |
Valvular
Insufficiency |
Pericarditis |
| Cardiac
Tamponade |
Femoral
Artery |
Pleural
Effusion |
| Cerebrovascular |
Laceration
|
Respiratory
Depression |
|
Accident
|
Thrombus/Embolic |
Atrial
Lead |
| Myocardial
Infarction |
Event
|
Dislodgement
|
| Unintended
AV or |
Bowel
Obstruction |
Infected
IV Site |
|
Bundle
Branch
|
Brachial
Plexus Injury |
Hypotension |
|
Block
Requiring new
|
Pneumonia |
Bradycardia |
|
pacemaker
(1.0%)
|
Inferior
Vena |
Vasovagal
Reaction |
| Coronary
Sinus |
Cava
Clot
|
Thrombophlebitis |
|
Perforation
|
Deep Vein |
Temperature
Elevation |
| Bacterial
Endocarditis |
Thrombosis
|
Puncture
Site |
| Ventricular
Fibrillation |
Pericardial |
Hematoma
(1.6%)
|
| Coronary
Artery Spasm |
Effusion
(2.2%)
|
Phlebothrombosis |
|
|
High CPK |
Other complications of catheter ablation
procedures which have been reported in the literature include:
| Pulmonary
Embolism |
Nerve
or Blood |
Transient
Ischemic Attack |
| Thromboembolism |
Vessel
Injury
|
Infection |
There were 18 deaths reported among the
683 patients enrolled in the study. None of the 18 deaths were judged
by the investigators as definitely related to the use of the Medtronic
RFCA System, but two deaths were judged as possibly device related
and a third as possibly procedure related.
Device related complications were reported
in 30 of the 683 patients (4%) including unintentional AV block
in 13 patients (1.9%), pericardial effusion in 8 patients (1.2%),
and cardiac tamponade in 3 patients (0.4%).
return to top
RF
Enhancr
Intended Use
The RF Enhancr catheter is intended for use with the Medtronic
RF power generator to deliver RF energy for intracardiac ablation
of accessory atrioventricular (AV) conduction pathways associated
with tachycardia for the treatment of AV nodal re-entrant tachycardia
and for creation of complete AV block in patients with a difficult
to control ventricular response to an atrial arrhythmia.
Contraindications
The use of this device is contraindicated in patients with active
systemic infection.
The transseptal approach is contraindicated in patients with left
atrial thrombus or myxoma, or interatrial baffle or patch. The retrograde
transaortic approach is contraindicated in patients with aortic
valve replacement.
Warnings and Precautions
General
- Related product literature Do not attempt to operate
the Medtronic ablation system or connect the ablation catheter
to a Medtronic RF power generator prior to completely reading
and understanding the Medtronic ablation system technical manual
and the ablation catheter instructions for use.
- System compatibility Use the catheter with only
a Medtronic RF power generator and accessories. The safety and
use with other RF power generators or accessories has not been
tested. Use only Medtronic cables.
- Expert users The catheter should be used only
by or under the supervision of physicians well trained in electrophysiology,
including the placement and use of intracardiac electrode catheters,
and experienced in performing RF catheter ablation procedures.
- Necessary environment Cardiac ablation procedures
should be performed only in a fully equipped electrophysiology
laboratory.
Ablation therapy hazards
- Serious adverse events A number of serious adverse
events have been documented for catheter ablation procedures,
including pulmonary embolism; myocardial infarction; cerebrovascular
accident; cardiac damage, perforation, and tamponade; perforation
of the vasculature; and death. See the Adverse events
section for additional potential adverse events.
- Left-sided ablation procedures Patients undergoing
left-sided ablation procedures should be closely monitored during
the postablation period for clinical manifestations of infarction.
- Distal pair electrode spacing of >2 mm Catheters
with distal pair electrode spacing greater than 2 mm should not
be used in the ablation of septal accessory pathways or in the
treatment of AV nodal re-entrant tachycardia because of the potential
for creating inadvertent complete AV block.
- Catheter manipulation and placement Provide adequate
fluoroscopic visualization during catheter manipulation and placement.
During a transaortic approach, avoid placement of the ablation
catheter within the coronary vasculature. Do not use excessive
force to advance or withdraw the catheter when resistance is encountered.
Catheter placement and RF power application within a coronary
artery have been associated with myocardial infarction and death.
- X-ray and fluoroscopic exposure Due to the x-ray
beam intensity and the duration of the fluoroscopic imaging during
ablation procedures, patients and laboratory staff may be subjected
to acute radiation injury and increased risk for somatic and genetic
effects. The long-term effects of protracted fluoroscopy have
not been established.
- Mimimize x-ray exposure.
- Carefully consider the use of the device in pregnant women
and prepubescent children.
- AV conduction Closely monitor AV conduction during
RF energy delivery in patients undergoing AV node modification
or septal accessory pathway ablation. These patients may be at
risk for complete atrioventricular (AV) block. Immediately terminate
energy delivery if partial or complete AV block is noted.
- Leakage current Use only isolated amplifiers,
pacing equipment, and ECG equipment (IEC 601-1 Type CF equipment,
or equivalent) or patient injury or death may occur. Leakage current
from any
connected device to the patient must not exceed 10 micro Amps
(mA)
under any circumstances.
- Catheter removal See your Medtronic ablation system
technical manual for information concerning catheter removal following
generator shutdown.
- Long-term risk The long-term risks of lesions
created by RF ablation have not been established. In particular,
any long-term effects of lesions in proximity to the specialized
conduction system or coronary vasculature are unknown. Furthermore,
the risk/benefit in asymptomatic patients has not been studied.
Catheter storage and handling
- Storage conditions Optimal storage conditions
for the catheter are these: 10-50 °C, up to 95% relative humidity,
noncondensing.
- Inspecting the sterile package Inspect the sterile
packaging and catheter prior to use. If either shows damage, do
not use the catheter; contact your local Medtronic representative.
- Single use only Do not resterilize or reuse Medtronic
catheters, which are disposable and intended for single use only.
Reuse may result in loss of proper electrical and mechanical function
and could cause patient injury and/or the communication of infectious
diseases from one patient to another.
Catheter handling and care
Do not excessively bend or kink the catheter. Excessive
bending or kinking may damage internal electrode wires and/or
distal tip shaping capabilities.
Do not allow moisture onto the connectors on the catheter,
Medtronic RF power generator, or cables. If the connectors get
wet, the system may not function correctly.
Do not wipe the catheter with organic solvents, such
as alcohol, to maintain optimal patient safety and catheter
electrode integrity.
Dispose of the catheter according to local environmental
requirements.
Adverse Events
One or more adverse events were reported in 67 of the 683 patients
(a total of 88 adverse events) during the clinical study of the
Medtronic ablation system. The following adverse events are listed
in descending order
according to their clinical significance as determined by their
severity and frequency (<1% unless otherwise indicated).
| Death |
Valvular
Insufficiency |
Pericarditis |
| Cardiac
Tamponade |
Femoral
Artery |
Pleural
Effusion |
| Cerebrovascular |
Laceration
|
Respiratory
Depression |
|
Accident
|
Thrombus/Embolic |
Atrial
Lead |
| Myocardial
Infarction |
Event
|
Dislodgement
|
| Unintended
AV or |
Bowel
Obstruction |
Infected
IV Site |
|
Bundle
Branch
|
Brachial
Plexus Injury |
Hypotension |
|
Block
Requiring new
|
Pneumonia |
Bradycardia |
|
pacemaker
(1.0%)
|
Inferior
Vena |
Vasovagal
Reaction |
| Coronary
Sinus |
Cava
Clot
|
Thrombophlebitis |
|
Perforation
|
Deep Vein |
Temperature
Elevation |
| Bacterial
Endocarditis |
Thrombosis
|
Puncture
Site |
| Ventricular
Fibrillation |
Pericardial |
Hematoma
(1.6%)
|
| Coronary
Artery Spasm |
Effusion
(2.2%)
|
Phlebothrombosis |
|
|
High CPK |
Other complications of catheter ablation procedures which have
been reported in the literature include:
| Pulmonary
Embolism |
Nerve
or Blood |
Transient
Ischemic Attack |
| Thromboembolism |
Vessel
Injury
|
Infection |
There were 18 deaths reported among the 683 patients enrolled in
the study. None of the 18 deaths were judged by the investigators
as definitely related to the use of the RFCA System, but two deaths
were judged as possibly device related and a third as possibly procedure
related. Device related complications were reported in 30 of the
683 patients (4%) including unintentional AV block in 13 patients
(1.9%), pericardial effusion in 8 patients (1.2%), and cardiac tamponade
in 3 patients (0.4%).
return to top
RF
Marinr® MC
Intended Use
The RF Marinr Conductr catheter is intended for use with the Atakr
RF Power Generator to deliver RF energy for intracardiac radiofrequency
ablation of accessory atrioventricular (AV) conduction pathways
associated with tachycardia, for the treatment of AV nodal re-entrant
tachycardia, and for creation of
complete AV block in patients with a difficult to control ventricular
response to an atrial arrhythmia.
Contraindications
The use of this device is contraindicated in patients with active
systemic infection. The transseptal approach is contraindicated
in patients with left atrial thrombus or myxoma, or interatrial
baffle or patch. The retrograde transaortic approach is contraindicated
in patients with aortic valve replacement.
Warnings
- Performing ablation from within the coronary sinus and its venous
branches with a 5 French RF Marinr Catheter may present additional
safety risks because of the increased likelihood of injurying
adjacent coronary arterial structures.
- Catheter ablation procedures present the potential for significant
x-ray exposure, which can result in acute radiation injury
as well as increased risk for somatic and genetic effects, to
both patients and laboratory staff due to the x-ray beam intensity
and duration of the fluoroscopic imaging. Catheter ablation should
only be performed after adequate attention has been given to the
potential radiation exposure associated with the procedure, and
steps taken to minimize this exposure. Careful consideration must
therefore be given to the use of the device in pregnant women.
- During a transaortic approach, adequate fluoroscopic visualization
is necessary to avoid placement of the ablation catheter within
the coronary vasculature. Catheter placement and RF power application
within a coronary artery has been associated with myocardial infarction
and death.
- Catheter materials are not compatible with magnetic resonance
imaging (MRI).
- Perforation of the vasculature is an inherent risk of
any catheter placement. A number of serious adverse events have
been documented for catheter ablation procedures including pulmonary
embolism, myocardial infarction, stroke, cardiac tamponade, and
death. See Adverse Events, for additional potential complications.
- Patients undergoing left-sided ablation procedures should
be closely monitored during the post-ablation period for clinical
manifestations of infarction.
- Catheters with distal pair electrode spacing greater than
2 mm should not be used in the ablation of septal accessory
pathways or in the treatment of AV nodal re-entrant tachycardia
because of the potential for creating inadvertent complete AV
block.
- Implantable pacemakers and implantable cardioverter/defibrillators
(ICDs) may be adversely affected by radiofrequency (RF) current.
It is important to: have temporary external sources of pacing
and defibrillation available during ablation; exercise extreme
caution during ablation when in close proximity to atrial or ventricular
permanent leads; and, perform complete implantable device analysis
for all patients after ablation.
- Implanted cardioverter/defibrillators should be deactivated
during delivery of RF power.
- Patients undergoing AV node modification or septal accessory
pathway ablation are at risk for complete AV block. Permanent
pacing was required in 1.6% (2/128) of septal accessory pathway,
3.1% (1/32) of left posterior pathways, and 1.7% (4/238) of AV
nodal modification patients who experienced inadvertent partial
or complete AV block during the study. Closely monitor AV conduction
during RF energy delivery. Immediately terminate energy delivery
if partial or complete AV block is noted.
- The Atakr RF Power Generator is capable of delivering significant
RF power. Do not touch the distal tip of the CardioRhythm
RF catheter and the dispersive electrode at the same time especially
while operating the Atakr RF Power Generator or operator injury
may occur.
- Use only isolated amplifiers, pacing equipment, and ECG equipment
(IEC 601-1 Type CF equipment, or equivalent) or patient injury
or death may occur. Leakage current from any connected device
to the patient must not exceed 10 micro Amps (µA) under
any circumstances.
- The CardioRhythm RF catheters are disposable and intended
for single use only. Do not resterilize or reuse. This may
result in loss of proper electrical and mechanical function and
could cause patient injury.
Precautions
- Due to the smaller diameter of the 5 French RF Marinr Catheter,
lesion dimensions may be smaller as compared to using a 7 or 8
French ablation system at similar power outputs.
- During in vitro testing with a 5 French RF Marinr Catheter under
no flow conditions, this catheter demonstrated an increased frequency
of temperature shutdowns.
- Do not attempt to operate the Atakr Ablation System or connect
the catheter to the Atakr RF Power Generator prior to completely
reading and understanding the Atakr Ablation System Technical
Manual and the RF Conductr MC catheter Instructions For Use.
- Cardiac ablation procedures should be performed only by appropriately
trained personnel in a fully-equipped electrophysiology
laboratory.
- This catheter should be used only by or under the supervision
of physicians well trained in electrophysiology, including the
placement and use of intracardiac electrode catheters and experienced
in performing radiofrequency catheter ablation procedures.
- The long-term risks of protracted fluoroscopy have not
been established. Careful consideration must therefore be given
for the use of the device in prepubescent children.
- The long-term risks of lesions created by RF ablation have
not been established. In particular, any long-term effects
of lesions in proximity to the specialized conduction system or
coronary vasculature are unknown. Furthermore, the risk/benefit
in asymptomatic patients has not been studied.
- The CardioRhythm RF catheters are intended for use only with
the Atakr RF Power Generator and accessories. The safety and
use with other RF power generators or accessories has not been
tested. Use only
Medtronic CardioRhythm cables supplied with the Atakr Ablation
System.
- Excessive bending or kinking of catheter may damage internal
electrode wires and/or distal tip shaping capabilities.
- Do not allow moisture into connectors on the CardioRhythm
RF catheter, Atakr RF Power Generator, or cables. If the connectors
get wet, the system may not function correctly.
- The sterile packaging and catheter should be inspected prior
to use. If damaged, do not use; contact your local Medtronic representative.
- Careful catheter manipulation must be performed in order to
avoid cardiac damage, perforation or tamponade. Catheter advancement
should be performed under fluoroscopic guidance. Do not
use excessive force to advance or withdraw the catheter when resistance
is encountered.
- See Atakr Technical Manual (Section 9.10) for information concerning
catheter removal following generator shutdown.
- Do not deliver DC energy through the Atakr Ablation System
or the CardioRhythm RF catheter. Neither the Atakr Ablation System
nor the CardioRhythm RF catheter were designed to deliver DC energy
and no
testing has been performed.
- Verify that RF catheter tip electrode is not in contact with
electrodes of other catheters in the heart or heating of the
other electrodes may occur.
- The catheter impedance display should be monitored during
energy delivery. If a sudden impedance rise is observed during
the ablation procedure, energy delivery should be discontinued
and the catheter tip
examined for coagulum, and coagulum removed if present.
- Do not begin an ablation procedure if the LOW BATTERY
display is illuminated on the Atakr RF Power Generator. Replace
battery.
- To maintain optimal patient safety and catheter electrode integrity
do not wipe this catheter with organic solvents, such as
alcohol.
- Bench testing of CardioRhythm RF catheters demonstrated that
they could withstand twenty-five RF energy deliveries without
any reduction in performance. During the clinical trial as many
as 73 energy deliveries were made with a single catheter, although
the average number was about seven.
Clinical Summary
Please refer to the Atakr Ablation System Technical Manual for the
Clinical Background.
Adverse Events
One or more adverse events were reported in 67 of the 683 patients
(a total of 88 adverse events) during the clinical study of the
Atakr Ablation System. The following adverse events are listed in
descending order according to their clinical significance as determined
by their severity and frequency (< 1% unless otherwise indicated).
| Death |
Valvular
Insufficiency |
Pericarditis |
| Cardiac
Tamponade |
Femoral
Artery |
Pleural
Effusion |
| Cerebrovascular |
Laceration
|
Respiratory
Depression |
|
Accident
|
Thrombus/Embolic |
Atrial
Lead |
| Myocardial
Infarction |
Event
|
Dislodgement
|
| Unintended
AV or |
Bowel
Obstruction |
Infected
IV Site |
|
Bundle
Branch
|
Brachial
Plexus Injury |
Hypotension |
|
Block
Requiring new
|
Pneumonia |
Bradycardia |
|
pacemaker
(1.0%)
|
Inferior
Vena |
Vasovagal
Reaction |
| Coronary
Sinus |
Cava
Clot
|
Thrombophlebitis |
|
Perforation
|
Deep Vein |
Temperature
Elevation |
| Bacterial
Endocarditis |
Thrombosis
|
Puncture
Site |
| Ventricular
Fibrillation |
Pericardial |
Hematoma
(1.6%)
|
| Coronary
Artery Spasm |
Effusion
(2.2%)
|
Phlebothrombosis |
|
|
High CPK |
Other complications of catheter ablation
procedures which have been reported in the literature include:
| Pulmonary
Embolism |
Nerve
or Blood |
Transient
Ischemic Attack |
| Thromboembolism |
Vessel
Injury
|
Infection |
There were 18 deaths reported among the
683 patients enrolled in the study. None of the 18 deaths were judged
by the investigators as definitely related to the use of the Atakr
RFCA System, but two deaths were judged as possibly device related
and a third as possibly procedure related.
Device related complications were reported
in 30 of the 683 patients (4%) including unintentional AV block
in 13 patients (1.9%), pericardial effusion in 8 patients (1.2%),
and cardiac tamponade in 3 patients (0.4%).
return to top
7F
Marinr ®
Indications for use
The Medtronic CardioRhythm Marinr Catheter is intended for use in
diagnostic electrophysiologic procedures. These catheters are designed
for recording intracardiac electrograms and temporary pacing associated
with electrophysiology studies.
Contraindications
There are no known contraindications for these catheters.
Warnings
- This device is for single use only. Do not resterilize and
reuse the catheter. This may result in loss of proper electrical
and mechanical function.
- This catheter should be used only by or under the supervision
of physicians well trained in electrophysiology, including the
placement and use of intracardiac electrode catheters.
- Electrical recording or stimulation equipment must be front
end isolated, or have an isolated patient cable. Current
leakage from any connected electrical equipment must not exceed
10 µA for intracardiac electrodes.
- Perforation of the vasculature is an inherent risk of any catheter
placement. Additional potential complications are those attending
any intracardiac catheterization procedure including, but not
limited to: cardiac tamponade, thromboembolic episodes, hematoma,
pneumothorax, local or systemic infection and death.
- Use only connecting cables supplied by Medtronic CardioRhythm
or patient injury, operator injury, or equipment damage may occur.
- Do not allow moisture into connectors of the Marinr Catheter
or connecting cables. If the connectors get wet, the system may
not function correctly.
- US LAW DOES NOT PERMIT THE USE OF THIS CATHETER FOR INTRACARDIAC
ABLATION. DO NOT USE FOR DC ABLATION.
Precautions
- To maintain optimal patient safety and catheter electrode integrity
do not wipe this catheter with organic solvents, such as alcohol.
- Excessive bending or kinking of catheter may damage internal
electrode wires and/or distal tip shaping capabilities.
- Catheter materials are not compatible with magnetic resonance
imaging (MRI).
return to top
Localisa®
Indications for Use
The Medtronic Localisa® Intracardiac Navigation System is
intended for real-time display of electrode positions in the heart
in relation to points taken and stored of previous electrode positions.
These positions can be displayed simultaneously to provide a simple
chamber geometry map of the heart.
The Localisa® Intracardiac Navigation System is indicated
for any procedure in which electrodes need to be navigated and/or
located within the heart. The Localisa® Intracardiac Navigation
System, when used with data from other devices such as an RF power
generator or an EP recording system, can generate ablation data
maps, cardiac electrical activation maps, cardiac electrical propagation
maps, or cardiac electrical potential maps.
General Cautions and Warnings
The CPU and monitor (non-medical equipment) should not be connected
directly to a wall outlet. Connecting this equipment directly to
a wall outlet may result in electrical safety hazards.
Accessory equipment connected to the analog and digital interfaces
must be certified according to the respective IEC standards (e.g.
IEC 950 for data processing equipment and IEC 601-1-1 for medical
equipment). Furthermore, all configurations shall comply with the
system standard IEC 601-1-1. Any person who connects additional
equipment to the signal input part or signal output part configures
a medical system, and is therefore responsible for ensuring that
the system complies with the requirements of the system standard
IEC 601-1-1. If in doubt, consult the technical service department
or your local representative.
Do not disconnect the LCD power or video connections and simultaneously
touch the patient.
Do not use the Localisa System in an anesthetically flammable environment.
The Localisa System has been successfully tested against the requirements
of IEC 60601-1, but the possibility remains that RF interface could
hamper system operation or that RF interference from the Localisa
System could hamper the operation of other nearby electrical devices.
If you suspect either of these conditions, try moving the conflicting
equipment farther apart, separating the equipment with an RF barrier,
or discontinue the use of the Localisa System.
Cautions
The Localisa System should be used only as an adjunct for intracardiac
navigation. The system is not a replacement for the physician's
knowledge, expertise, or judgment.
The Localisa System should be used only by qualified medical professionals
who have been thoroughly trained and are experienced in its use.
The Localisa System is designed for use with only the Medtronic
Atakr II RF Power Generator. No other RF power generator should
be used.
Skin electrodes used to generate the Localisa Coordinate system
Voltage Gradient must be placed orthogonally, mutually perpendicular,
across the patient's body, targeting the heart as the centroid of
the orthogonal system field.
The Localisa System is designed for use with the Medtronic 6416
Temporary Pacing Lead as the intracardiac reference for voltage
measurements. The lead must be properly fixated into the wall of
the heart chamber in which the procedure will be performed.
If an ablation is performed, there is a possibility that positional
information may be invalid during the ablation and up to 30 seconds
after the end of the ablation. If the Atakr II RF power Generator
is connected to the Localisa Workstation via an RS232 serial link,
the System Software will prompt the operator that an ablation is
in progress. The System will not allow collections of data points
for 30 seconds after an ablation.
The Power Supply Enclosure Unit weighs approximately 95 pounds
(43kg). Medtronic recommends that two people lift and place the
unit. Use both hands and lift the unit from the bottom.
The Localisa System components contain no user-repairable parts.
For repair or replacement of any part of the system, contact technical
support.
return to top
Brockenbrough
Needle
The Medtronic Brockenbrough curved needle
is made up of an outer cannula and an inner stylet. The outer cannula
is made of flexible thin-walled tubing. The inner stylet is solid,
much stiffer and closely fitting, whose sharpened end protrudes
2-3 mm beyond the tip of the cannula.
Indications for use
The Brockenbrough curved needle is used in conjunction with transseptal
catheters to puncture the atrial septum to allow conducting a left
heart catheterization procedure through the right atrium. The Brockenbrough
curved needle is intended for single use only.
Contraindications
The use of the Brockenbrough curved needle is contraindicated in
patients with the following conditions.
- distorted anatomy due to congenital heart disease
- significant chest or spine deformity
- the inability to lie flat
- ongoing anticoagulation
- marked atrial enlargement
- left atrial thrombus or tumor
- dilated aortic root
- previous patch repair of the interatrial septum
Warnings and precautions
- The Brockenbrough curved needle is designed for use by physicians
engaged in the practice of specialized invasive cardiology techniques.
Use of this needle should be restricted to those specialists trained
to perform transseptal procedures to minimize the complications
associated with this procedure.
- The Brockenbrough curved needle must be sharp to minimize the
risk of puncturing the left atrial wall just after crossing the
septum.
- Use continuous pressure monitoring and repeated biplane fluoroscopy
of the tip during any positioning, to minimize the risk of advancing
the Brockenbrough curved needle into an undesired location and
the subsequent bleeding that might occur.
- Extreme caution should be used in patients with a small left
atrium, to avoid puncture of the left atrial wall.
- Never advance the Brockenbrough curved needle until the catheter
is correctly positioned on the atrial septum to minimize the risk
of advancing the needle into an undesired location.
- Never advance a dilator, sheath or catheter until the Brockenbrough
curved needle has clearly entered the left atrial cavity, as confirmed
by a distinct left atrial pressure record and the tip of the needle
is in the left atrium on fluoroscopy.
- If a dilator, sheath or catheter are inadvertently advanced
into the pericardial space, do not remove them until the patient
is in surgery.
- Do not reuse the Brockenbrough curved needle. The Brockenbrough
curved needle must be discarded after one procedure. Its structural
integrity and/or function can be impaired through reuse or cleaning.
All parts are extremely difficult to clean after being exposed
to biological materials, and adverse patient reactions can result
from reuse.
- The Brockenbrough curved needle has been gamma sterilized prior
to shipment. Carefully inspect the package before opening. Damaged
packages should be returned to Medtronic. The Brockenbrough curved
needle cannot be resterilized.
- Do not use the product after its expiration date.
- At no time should the introducer or guide wire be advanced or
withdrawn when resistance is met without first determining the
cause by fluoroscopy and taking remedial action.
- Ensure that the catheter is appropriate for the Brockenbrough
curved needle being used.
- After use, the Brockenbrough curved needle may be a potential
biohazard. Handle and dispose of it in accordance with accepted
medical practice and applicable local, state and federal laws
and regulations.
Adverse events
In addition to the complications associated with any cardiac catheterization,
the following can occur during a transseptal catheterization:
- puncture of the atrial free wall
- puncture of the aorta
- puncture of the inferior vena cava
- puncture of the coronary sinus
- tamponade
- arterial embolism from thrombus at the puncture site
- atrial arrhythmias
- residual atrial septal defects
return to top
RF
Contactr
Intended use
The RF Contactr catheter is intended for use with a Medtronic
RF power generator to deliver RF energy for intracardiac radiofrequency
ablation of accessory atrioventricular (AV) conduction pathways
associated with tachycardia for the treatment of AV nodal re-entrant
tachycardia and for creation of complete AV block in patients with
a difficult to control ventricular response to an atrial arrhythmia.
Contraindications
The use of this device is contraindicated in patients with active
systemic infection. The transseptal approach is contraindicated
in patients with left atrial thrombus or myxoma, or interatrial
baffle or patch. The retrograde transaortic approach is contraindicated
in patients with aortic valve replacement.
Warnings and precautions
General
Related product literature Do not
attempt to operate the Medtronic ablation system or connect the
ablation catheter to a Medtronic RF power generator prior to completely
reading and understanding the Medtronic ablation system technical
manual and the ablation catheter technical manual.
System compatibility Use the catheter
with only a Medtronic RF power generator and accessories. The safety
and use with other RF power generators or accessories has not been
tested. Use only Medtronic cables.
Users The catheter should be used
only by or under the supervision of physicians trained in electrophysiology.
This training includes the placement and use of intracardiac electrode
catheters and experience in performing RF catheter ablation procedures.
Necessary environment Cardiac ablation
procedures should be performed only in a fully equipped electrophysiology
laboratory.
Ablation therapy hazards Left-sided
ablation procedures Patients undergoing left-sided ablation
procedures should be closely monitored during the post-ablation
period for clinical manifestations of infarction.
Distal pair electrode spacing of >2
mm Catheters with distal pair electrode spacing greater than
2 mm should not be used in the ablation of septal accessory pathways
or in the treatment of AV nodal re-entrant tachycardia because of
the potential for creating inadvertent complete AV block.
Catheter manipulation and placement
Provide adequate fluoroscopic visualization during catheter manipulation
and placement. During a transaortic approach, avoid placement of
the ablation catheter within the coronary vasculature. Do not use
excessive force to advance or withdraw the catheter when resistance
is encountered. Catheter placement and RF power application within
a coronary artery have been associated with myocardial infarction
and death.
X-ray and fluoroscopic exposure
Due to the x-ray beam intensity and the duration of the fluoroscopic
imaging during ablation procedures, patients and laboratory staff
may be subjected to acute radiation injury and increased risk for
somatic and genetic effects. Because the long-term effects of protracted
fluoroscopy have not yet been established, please take the following
precautions.
- Minimize x-ray exposure.
- Carefully consider the use of x-ray and fluorocopy in pregnant
women and prepubescent children.
AV conduction Closely monitor
AV conduction during RF energy delivery in patients undergoing AV
node modification or septal accessory pathway ablation. These patients
may be at risk for complete atrioventricular (AV) block. Immediately
terminate energy delivery if partial or complete AV block is noted.
Leakage current Use only isolated
amplifiers, pacing equipment, and ECG equipment (IEC 601-1 Type
CF equipment) or patient injury or death may occur. Leakage current
from any connected device to the patient must not exceed 10 “A under
any circumstances.
Catheter removal Before removing
the catheter, ensure that the catheter tip is not curved or deflected.
Patient injury may occur.
Long-term risk of lesions The
long-term risk of lesions created by RF ablation have not been established.
In particular, any long-term effects of lesions in proximity to
the specialized conduction system or coronary vasculature are unknown.
Furthermore, the risk/benefit in asymptomatic patients has not been
studied.
Catheter storage and handling Storage
conditions Optimal storage conditions for the catheter are:
10-50 ˇC (50-122 ˇF), up to 95% relative humidity, noncondensing.
Inspecting the sterile package Inspect the sterile package and
catheter prior to use. If either shows damage, do not use the catheter
and contact your local Medtronic representative.
Single use only Do not resterilize
or reuse Medtronic catheters. Medtronic catheters are disposable
and intended for single use only. Reuse may result in loss of proper
electrical and mechanical function and could cause patient injury
and/or the communication of infectious diseases from one patient
to another.
Catheter handling and care
Follow the precautions below when handling the catheter.
- Do not excessively bend or kink the catheter. Excessive bending
or kinking may damage internal electrode wires and/or distal tip
shaping capabilities.
- Do not allow moisture onto the connectors on the catheter, Medtronic
RF power generator, or cables. If the connectors get wet, the
system may not function correctly.
- Do not wipe the catheter with organic solvents (i.e. alcohol).
Using organic solvents on the catheter degrades the catheter's
electrode integrity, and thereby diminishes patient safety.
- Dispose of the catheter according to local environmental requirements.
return to top
Atakr®
II
The Atakr II RF Generator provides high power output with the integral
safety of closed-loop temperature control using the embedded thermocouples
in Medtronic RF ablation catheters.
Features:
- 60 Watt output
- Closed-loop temperature control
- Operates in power or temperature mode
- LEM feature provides tip to tissue contact
- Adjustable impedance parameters
- Adjustable power and temperature limits for safety
- Remote control electronic foot pedal
- Line powered
The System comes equipped with the following:
1 60 Watt RF Generator
1 Electronic foot pedal
1 Atakr to RF catheter cable
1 Dispersive electrode cable
1 Atakr to EGM cable
|